Direct estimation of central aortic pressure from measured or quantified mean and diastolic brachial blood pressure: agreement with invasive records

Background: Recently it has been proposed a new approach to estimate aortic systolic blood pressure (aoSBP) without the need for specific devices, operator-dependent techniques and/or complex wave propagation models/algorithms. The approach proposes aoSBP can be quantified from brachial diastolic an...

Celý popis

Uloženo v:
Podrobná bibliografie
Hlavní autor: Bia, Daniel (author)
Další autoři: Salazar, Federico (author), Cinca, Luis (author), Gutiérrez, Marcos (author), Facta, Álvaro (author), Zócalo, Yanina (author), Díaz, Alejandro (author)
Médium: article
Jazyk:angličtina
Vydáno: 2023
Témata:
On-line přístup:https://hdl.handle.net/20.500.12008/52088
Tagy: Přidat tag
Žádné tagy, Buďte první, kdo vytvoří štítek k tomuto záznamu!
Popis
Shrnutí:Background: Recently it has been proposed a new approach to estimate aortic systolic blood pressure (aoSBP) without the need for specific devices, operator-dependent techniques and/or complex wave propagation models/algorithms. The approach proposes aoSBP can be quantified from brachial diastolic and mean blood pressure (bDBP, bMBP) as: aoSBP = bMBP2/bDBP. It remains to be assessed to what extent the method and/or equation used to obtain the bMBP levels considered in aoSBP calculation may affect the estimated aoSBP, and consequently the agreement with aoSBP invasively recorded. Methods: Brachial and aortic pressure were simultaneously obtained invasively (catheterization) and non-invasively (brachial oscillometry) in 89 subjects. aoSBP was quantified in seven different ways, using measured (oscillometry-derived) and calculated (six equations) mean blood pressure (MBP) levels. The agreement between invasive and estimated aoSBP was analyzed (Concordance correlation coefficient; Bland-Altman Test). Conclusions: The ability of the equation “aoSBP = MBP2/DBP” to (accurately) estimate (error <5 mmHg) invasive aoSBP depends on the method and equation considered to determine bMBP, and on the aoSBP levels (proportional error). Oscillometric bMBP and/or approaches that consider adjustments for heart rate or a form factor ∼40% (instead of the usual 33%) would be the best way to obtain the bMBP levels to be used to calculate aoSBP.